REFERENCE
CLASS CHAMBERS
Reference
Class Chambers are designed to be used by calibration laboratories that
employ reference grade electrometers and bias supplies.
•
Click Chamber Images for Energy Dependence
Charts and Schematic Drawings
• Click
here for Nominal Chamber Volumes and Sensitivity Chart below
•
Click here for Disconnect Cable Length
and Electrometer Termination
RC6
General Purpose X-Ray Measurements
The RC6 is an unsealed cylindrical 6 cm3, general-purpose ion
chamber, designed for general purpose x-ray measurements. It
has excellent energy response throughout the diagnostic energy
range and with an appropriate build-up cap (optional), the energy
range can be extended to 60Co.
| Exposure
Rate Dependence: |
|
<5%
to 150 mGy/s, up to 4 Gy/s for 10 ms pulses |
| |
|
|
| Energy
Dependence: |
|
±5%,
30 keV to 1.33 MeV (with build-up material) |
| |
|
|
| Bias:
|
|
Nominal
+300 VDC (max +600 VDC) |
| |
|
|
| Electrical
Leakage: |
|
<5e-15
A with +300 VDC bias |
| |
|
|
| Construction: |
|
Guarded (3-terminal) concentric cylinder with polycarbonate
walls and conducting graphite interior coating. Nominal
6 cm3 active volume; 0.05 kg |
|
|
RC6M
Low Energy X-Ray Measurements
The RC6M is an unsealed parallel plate 6 cm3 ion chamber specifically
designed for low-energy x-ray measurements. It has a thin entrance
window and is suitable for x-ray energies in the range of 10 to
40 keV.
| Exposure
Rate Dependence: |
|
<5%
to 90 mGy/s |
| |
|
|
| Energy
Dependence: |
|
±5%,
10 keV to 40 keV |
| |
|
|
| Bias: |
|
Nominal
+300 VDC (max +600 VDC) |
| |
|
|
| Electrical
Leakage: |
|
<5e-15
A with +300 VDC bias |
| |
|
|
| Construction: |
|
Guarded
(3-terminal) parallel plate, with 0.7mg/cm2 metalized polyester
window; polyacetal exterior; nominal 6 cm3 active volume; 0.08 kg |
|
|
RC60
X-Ray Beam Measurements
The RC60 is an unsealed thin cross-section, parallel plate 60 cm3
ion chamber suitable for x-ray beam measurements with low to intermediate
dose rates. It has excellent energy response throughout the diagnostic
energy range and with the application of appropriate build-up
materials, the energy range can be extended to 60Co.
| Exposure
Rate Dependence: |
|
<5%
to 30 mGy/s |
| |
|
|
| Energy
Dependence: |
|
±5%,
30 keV to 1.33 MeV (with build-up material) |
| |
|
|
| Bias: |
|
Nominal
+300 VDC (max +600 VDC) |
| |
|
|
| Electrical
Leakage: |
|
<5e-15
A with +300 VDC bias |
| |
|
|
Construction:
|
|
Guarded
(3-terminal) parallel plate, with polycarbonate walls; conductive
graphite interior and exterior coating; nominal 60 cm3 active volume,
0.5 m low-noise triax cable; 0.13 kg.
|
| Note: The chamber assembly shown in the image depicts the chamber with the Radcal Model 8230M chamber support, optional, ordered separately. |
|
|
RC3CT
CT X-Ray Beam Measurements
The RC3CT is an unsealed cylindrical, 10 cm pencil ion chamber with a 3
cm3 volume. It is designed specifically for CT x-ray
beam measurements, either free-in-air or mounted in a head or
body phantom. It has excellent energy and partial volume
response along its entire 10 cm length.
| Exposure
Rate Dependence: |
|
<5%
to 350 mGy/s |
| |
|
|
| Energy
Dependence: |
|
±5%,
3-20 mm Al hvl |
| |
|
|
| Bias: |
|
Nominal
+300 VDC (max +600 VDC) |
| |
|
|
| Electrical
Leakage: |
|
<5e-15
A with +300 VDC bias |
| |
|
|
| Construction: |
|
Guarded
(3-terminal) 10 cm long cylindrical, with C552 air-equivalent
walls and electrode; polyacetal exterior
cap; nominal 3 cm3 active volume; 0.5 m, low-noise triax cable; 0.11
kg |
|
|
RC180
Low Dose Rate X-Ray Beam or Scatter Measurements
The RC180 is an unsealed parallel plate 180 cm3 ion chamber suitable
for low dose rate x-ray beam or scatter measurements. It has an
effective cross section of 100 cm3 and has excellent
energy response throughout the diagnostic energy range. With the
application of appropriate build-up materials, the energy range
can be extended to 60Co.
| Exposure
Rate Dependence: |
|
<5%
to 5 mGy/s |
| |
|
|
| Energy
Dependence: |
|
±5%,
33 keV to 1.33 MeV (with build-up material) |
| |
|
|
| Bias: |
|
Nominal
+300 VDC (max +600 VDC) |
| |
|
|
| Electrical
Leakage: |
|
<5e-15
A with +300 VDC bias |
| |
|
|
| Construction: |
|
Guarded
(3-terminal) parallel plate, with polycarbonate walls and
electrode; conducting graphite interior and exterior coating;
nominal 180 cm3 active volume; 0.11 kg |
|
|
RC1800
Low Dose Rate Radiation Applications
The RC1800 is an unsealed cylindrical 1800 cm3 ionization
chamber intended for low dose rate radiation applications
such as shielding leakage, low-level irradiator output and environmental
measurements.
| Exposure
Rate Dependence: |
|
<5% to 50 uGy/s, <10% to 160 uGy/s (bias +300 VDC) |
| |
|
|
| Energy
Dependence: |
|
±5%,
33 keV to 1.33 MeV |
| |
|
|
| Bias: |
|
Nominal
+300 VDC (max +600 VDC) |
| |
|
|
| Electrical
Leakage: |
|
<1e-14
A with +300 VDC bias |
| |
|
|
| Construction: |
|
Guarded
(3-terminal) parallel plate, polycarbonate walls and electrode;
conducting graphite interior and exterior coating; 1800 cm3 active volume; 0.54 kg |
|
|
Nominal Chamber Volumes and Sensitivity
(±10%)
| CHAMBER
MODEL |
Volume
cm3 |
C/mGy
20 ºC |
C/mGy
22 ºC |
C/R
20 ºC |
C/R
22 ºC |
| RC6 |
6 |
2.1E-10 |
2.1E-10 |
1.9E-09 |
1.9E-09 |
| RC6M |
6 |
2.1E-10 |
2.1E-10 |
1.9E-09 |
1.9E-09 |
| RC60 |
60 |
2.1E-09 |
2.1E-09 |
1.9E-08 |
1.9E-08 |
| RC3CT |
3 |
1.1E-10 |
1.1E-10 |
9.3E-10 |
9.3E-10 |
| RC180 |
180 |
6.4E-09 |
6.3E-09 |
5.6E-08 |
5.6E-08 |
| RC1800 |
1800 |
6.4E-08 |
6.3E-08 |
5.6E-07 |
5.6E-07 |
Disconnect
Cable and Electrometer Termination
|
|
 |