Upgrades for your Dynalyzer Products
Upgrade to the NEW Radcal Dynalyzer IIIU
- Dynalyzer II
- Dynalyzer III
- Dynalyzer IIIA
- Dynalyzer IIIACT HVU
Benefits :
- Measure the filament current of CT and HF generators by extending the measurement to DC. The older Dynalyzers did not measure filament current below about 20 Hz. This feature utilizes a Hall-effect current sensor to replace the transformer used previously. The bandwidth of the new filament-current measurement extends from DC to 20 KHz.
- Increased sensitivity for measuring anode current in fluoroscopic systems. In Fluoro mode the anode-current measurement is 20 times more sensitive.
- Wide-range power supply. Allows accurate measurements in the presence of power-line disturbances. With the new supply line-voltage sags to 90V have no effect.
- Re-Engineered signal-processing circuitry to reduce noise and drift. The instrument will remain in calibration longer.
- Improved layout and added protective circuits so the signal-processing circuitry better withstands tube arcing.
- Fill the High-voltage unit to the maximum rated pressure with sulfur hexafluoride insulating gas. This assures safe and accurate measurement at the maximum rated voltage.
- Helium leak test the High-voltage unit to find and fix slow leaks that can reduce the insulating-gas pressure to unsafe levels.
- Full factory calibration after testing the completed system for conformance to current specifications.
SENDING YOUR INSTRUMENT TO US
No Return Authorization # is required, but for a fast turn around time, please send a PO for the proper amount along with the completed Service Request Form.
Our address is 426 West Duarte Road, Monrovia, CA 91016, USA.
Contact Radcal Calibration and Repair Services at: 1-626-357-7921 ext 119
Please note: If you are experiencing difficulties with your Radcal instrument, please call or mail service@radcal.com first. Some difficulties may be resolved without having to send your unit to us.
One Business Day Guaranteed Response!