BROCHURE
KEY FEATURES
For Radiography, Fluoroscopy, Mammography, CT, and Dental
SPECIFICATIONS
 
 
  XLPRO SOFTWARE (optional) click chart
 
 
Data Display Examples
 
INCREASED PRODUCTIVITY - Easy use, quick setup, convenient portability-all in one compact system.
VERSATILITY AND ACCURACY - Basic and advanced measurement functions for Critical Acceptance Testing and QA consistency tests.
SENSOR SELECTION - ACCU-kV kV sensors and Invasive/Non-invasive mA/mAs sensors, for all your measurement needs.
NEW ! - Now calibrate Hologic Selenia W-Rh, Ag tube kVp's. And, be sure to use our proven 10X6-6M ion chamber for dose with all anode - filter tube combinations, current and future.

 

   
   
Multi-Purpose meter
9096
Dosimeter
2186
Dosimeter
2086
kV-mA-mAs meter
4086
FUNCTION
   
Dose Accumulate Hold
 
Dose Rate
Auto Dose
Last Dose
Max Dose Rate
Pulsed Cine w/ Auto Sense
Exposure Time
* kV (filtration correction)
* Flash HVL
* Beam filtration
mA / mAs
Optional XLPRO Software
 


     
Simultaneous Measurements
Measures kVp or kV Average or PPV, Time, mA/mAs
FFT kV
Accurate kVp measurements for AMX4+ and AMX 700 portables
Flash HVL 
Displays HVL and beam filtration for Diagnostic kV sensor range
mA/mAs
Invasive and Non-Invasive sensor options
Data Capture Software
Auto data capture, waveforms and templates
Plug and Play Sensors
Sensors are interchangeable with any Accu-kV and Accu-Pro
 

 

 
 
 
  FFT kV SENSORS with FLASH HVL (diagnostic range)
The Diagnostic range sensor accuracy is ± 1 kV or ± 1 % whichever
is largest. The mammo sensor has an Mo-Mo accuracy of ± 0.5 kV
and < ± 1 kV for other track-filters.

 


AVAILABLE MEASUREMENT FUNCTIONS: Built-in digital display and computer display with optional XLPRO software.
               
 
kV
 
Time
 
mA
 
Beam Quality
 
kV peak, kV average, kV
 
Seconds
 
mA
 
Filtration (Al and Cu equivalents)
 
practical for: pulse (single
 
Minutes
 
mAs
 
HVL
 
exposure), fluoro, and
 
Hours
 
   
dental