Wide Beam CT Ion Chambers
Reference Class Ion Chambers
|
| |
|
|
|
| |
ACCESSORIES - BP* PORTABLE TEST STAND |
|
|
|
|
|
| |
| |
Testing of Fluoroscopic and X-ray System |
|
| |
|
Entrance
Exposure Rates • Collimation • Horizontal Beam Measurements •
HVL • AEC • kVp • Fixed Tube Exposure and more...
Entrance Skin Exposure Rate |
 |

|
| When
measuring Entrance
Skin Exposure Rate, position aluminum attenuators
below the chamber. |
|
 |
| |
Collimation Test Tool • Slightly Above Cassette
|
 |
Position
the collimation test tool slightly above the cassette. A 14"x17"
cassette may be used when the test stand's vertical supports
are inserted into the outer notches of the top of the cross
pieces. Cassettes may be exchanged without disturbing the test
tool. |
|
 |
| |
| Collimation Test Tool • Well Above Cassette |
 |
Position
the collimation test tool well above the cassette. A 14"x17"
cassette may be used when the test stand's vertical supports
are inserted into the outer notches of the top of the cross
pieces. Cassettes may be exchanged without disturbing the test
tool. |
|
 |
| |
Entrance
Skin Exposure Rate
|
Under
Table Fluoro Unit

Use
the test stand to position aluminum attenuators above the
chamber when measuring Entrance Skin Exposure Rate for an
Under Table Fluoro Unit.
|
C-Arm
Fluoro Unit - Close Up

Use
the test stand to safely position aluminum attenuators above
the chamber when measuring Entrance Skin Exposure Rate for
a C-Arm Fluoro Unit.
|
Lateral
C-Arm Fluoro Unit

Use
the test stand to position aluminum attenuators when measuring
Entrance Skin Exposure Rate for a Lateral C-Arm Fluoro Unit.
|
Lateral
C-Arm Fluoro Unit - Close Up

Use
the test stand to position aluminum attenuators when measuring
Entrance Skin Exposure Rate for a Lateral C-Arm Fluoro Unit.
|
C-Arm
Fluoro Unit - Above Chamber

Use
the test stand to position aluminum attenuators when measuring
Entrance Skin Exposure Rate for a Lateral C-Arm Fluoro Unit.
|
C-Arm
Fluoro Unit - Below Chamber

Use
the test stand to position aluminum attenuators below the
chamber when measuring Entrance Skin Exposure Rate for a C-Arm
Fluoro Unit.
|
|
 |
| |