Measures Dose, Dose Rate, kV, Time and mA or mAs
DISCONTINUED
MODEL - See the "NEW ACCU-PRO™ System"
Specifications for reference only
Full
Computer Control and Data Capture or Manual Control
Waveforms
for kV, FFT kV, Relative Dose and mA
OVERVIEW OF FEATURES AND BENEFITS
| Get
Increased Productivity |
| •
Intuitive, quick setup. |
| •
Large, clear, simple display to minimize
misreads. |
| •
Several specialized choices for both dose and dose rate functions add utility. |
|
• Auto
(and manual) fluoroscopic pulse rate and dose per pulse operation. |
|
• Single
or multiple function measurements save time. |
| •
High sensitivity with accuracy for CR
readers with filtration and for mini C-arms. |
| •
Powerful XLPRO-USB software is user customizable
for productivity. |
| •
Two quickly accessible alkaline C cells
for low drain Micro Power Circuitry. |
|
• Test
buttons and startup self-diagnostics verifies instrument status. |
|
• Autoranging
for low or high dose exposures. |
| Get
Accuracy and Confidence from Best Available Technology |
| •
Ion chamber basic physics technology.
No beam quality approximation algorithms ever necessary.
Obtain x-ray machine acceptance grade accuracy. |
|
•
Re-engineered compact built-in case. Enjoy no compromise on
performance for reduced size. |
| •
Unique advanced FFT circuitry for auto-ranging
kVp measurements. Manufacturers' kVp accuracy
for portables such as the GE AMX4, AMX 700, AMX
800, as well as analog and digital mammography. |
| Get
Cost Effectiveness |
| •
Extra features at no extra cost. |
|
• Scatter
and leakage capability -FDA 88-8035. |
| •
Easy upgrades: we invented expand-as-you-grow-sensors
and firmware. |
| •
Responsive factory direct support. |
| •
Interchangeable plug and play sensors. |
|
• Adjustable
automatic power-down enhances battery life and readiness. |
|
• Backward
compatiblity with standard 10X5 series Radcal ion chambers. |
| •
Available extended warranty program. Annual
calibration always includes PM. |
| •
Radcal designed for durability: 30+ years
of know-how. |
Please
see below for details

Proven
Technology
The 9095 Multi-Purpose Analyzer combines over 25 years of proven
technological features of Radcal's:
-
- 9000 series
dosimeters
- Accu-kV
4080 series non-invasive kV instruments
- mAs and
more
| Portable
And Compact
One system for all your diagnostic and mammography measurement
needs. Users familiar with the operation of the 9000-series
radiation monitors will find the operation of the 9095 very
similar but with added convenience features and a wider dynamic
range. |
Undisputed
Measurement Results
Whether you plan to use the 9095 for compliance, acceptance
testing, service or QA measurements, the 9095 will give you
the confidence you need to make accurate measurements with
reliable results. |
Mammography
Setup |
Diagnostic
Setup |
Single
Cable with Dual Termination deploys easily |
|
MULTIPLE FEATURES AND MEASUREMENTS
FEATURES
Multiple
measurements in a single exposure.
Integral cable for easy deployment &
rapid connection to one of many ion chambers & Accu-kV
sensors.
Convenient carry case allows for easy expansion
and storage of all accessories.
Automatic Pressure and Temperature correction
and direct readout eliminates the need for user intervention.
Automatic Recognition of Chamber and Sensor
Identification, Calibration factors and Date. |
Interchangeable ion-chambers and kV sensors with
any 9095 Analyzer.
Intrinsic Filtration - kV compensation for
user-entered total filtration between 1.2 and 10 mm Al.
Long battery life and automatic shut down
after 15 minutes of inactivity.
kV sensor automatically provides a relative
dose output without an ion chamber.
Analog output - Calibrated Real-time kV output,
20mV/kV.
|
MEASUREMENTS
Measurement
Functions: Dose, Auto Dose,
Last Dose, Dose Rate, Max Dose Rate, Pulsed/Cine, kV Pulse,
kV Fluoro and kV Dental. A setup mode allows the user
to set time and radiation units, temperature and pressure
compensation modes, choose or
eliminate main-menu functions, and control other parameters.
Dose:
Total dose (no threshold) during diagnostic, mammography and
low-level leakage exposures.
Auto
Dose: Automatic "hands-off," total-dose
measurements such as repeatability, mA linearity, and timer
linearity procedures.
Last
Dose: Provides a snap shot of total accumulated dose
at any time during a continuous exposure.
Dose
Rate: For measurements such as dose rate in fluoro
procedures and shielding integrity. Update rate is adjusted
auto-
matically for best results.
Max
Dose Rate: Display holds the peak dose rate measured
during an exposure, such as in fluoroscopy.
Pulsed/Cine:
Provides dose per pulse for Cine and pulsed fluoroscopic machines.
Automatic triggering, user selectable delay and acquisition
intervals.
kV
Pulse: Displays kVp, dose, time and mAs depending
on sensors connected for radiography and mammography.
kV
Fluoro: Continuous display of kVp, dose rate and
mA depending on sensors connected for Fluoroscopy.
kV
Dental: Operates like kV Pulse. Dental mode allows
use of an increased user selectable triggering threshold to
exclude preheat pulses from the kV and duration measurement.
|
ION
CHAMBERS
THE
GOLD STANDARD IN ION CHAMBER and kV SENSOR TECHNOLOGY
Radcal’s Model 9095 radiation-monitoring system utilizes a
wide range of Radcal ion chambers. The 9095 recognizes the connected
sensor and retrieves calibration information from the sensor; there
are no user-controlled calibrations required.
The Ion Chambers feature:
Uniform
energy response • Wide dynamic range • Proven advanced
performance
10X9-6
The General Purpose, in Beam Chamber
The 10X9-6 chamber permits in-beam measurements as described
in “Routine Compliance Testing for Diagnostic x-ray
systems” (PB89-205215). This chamber is ideal when using
the CDRH geometry (the 10T5 test stand). |
|
10X9-6M
The Dedicated Mammography Chamber
Long before the Mammography Quality Assurance Standards Act,
Radcal designed what proved to be one of the finest low energy
chambers on the market. It’s extraordinary flat energy
response over the 10 - 40 keV has been documented in technical
papers and makes corrections unnecessary. Unlike solid state
detectors, the -6M’s response is not influenced by tube
target material or filtration. |
|
10X9-60
& 10X9-60E
The ‘Service’ Chambers
The dynamic range and thin profile of these chambers make
them ideal for virtually all X-ray service applications:
• Input Dose at the Image Intensifier
• High dose rates encountered in Fluoroscopy
• Cine, spot film devices & other special procedures
Additionally, the -60E(extended) chamber’s increased
sensitivity at lower energies turns the chamber into a “Universal”
detector, covering mammography through R&F and beyond.
The 8231 holder is recommended. |
|
10X9-3CT
The Chamber for Computed Tomography Dose Index (CTDI)
When used in conjunction with Radcal’s 20CT6 (head phantom)
and Radcal’s 20CT14 (body phantom) the 10X9-3CT permits
evaluation of the radiation output as prescribed by HHS publication
FDA 88-8035. It’s excellent energy, partial volume response
and position uniformity has been well documented. |
|
10X9-180
The Leakage and Low Level Measurements Chamber
The 10X9-180 ion chamber is designed for leakage measurements
as described in “Regulations for the Administration
and Enforcement of The Radiation Control for Health and Safety
Act of 1968” (HHS Publication FDA 88-8035), and provides
the required effective cross-section of 100 cm2.
|
|
10X9-1800
The Radiation Protection Chamber
The 10X9-1800 ionization chamber is intended for low-level
radiation measurements such as shielding leakage, irradiator
and environmental. Unlike typical survey meters, the 10X9-1800
chamber offers improved accuracy over a wider dynamic range. |
|
CHAMBER OVERVIEW
| Chambers |
10x9-6 |
10x9-6M |
10x9-60 |
10x9-60E |
| Min
Rate |
1
µR/s
|
10
nGy/s |
1
µR/s |
10
nGy/s |
100
nR/s. |
1.nGy/s |
same |
| Max
Rate |
22
R/s |
190
mGy/s |
10
R/s |
90 mGy/s |
2.2
R/s |
19
mGy/s |
same |
| Min
Dose |
10
µR |
100
nGy |
10
µR |
100
nGy |
1
µR |
10
nGy |
same |
| Max
Dose |
79
kR |
.0.7
kGy |
79
kR |
0.7
kGy |
7.9
kR. |
.70
Gy |
same |
| Cine
Specifications |
0.1
µR/f - > 1 R/f
1 nGy/f - >10
mGy/f |
N/A |
0.01 µR/f - > 100 mR/f
0.1 nGy/f - >1.0 mGy/f |
same |
| Calibration
Accuracy |
±4%
using X-rays @ 60kVp and 2.8 mm AL HVL |
±4%
using X-rays @ 20kVp and 0.26 mm AL HVL |
(-60)
±4% using X-rays @ 150kVp and 10.2 mm AL HVL
(-60E)
±4%
using X-rays @ 50kVp and 0.88 mm AL HVL |
| Exposure
Rate Dependence |
±5%,
0.4 mR/s to 80 R/s, up to 500 R/s for 50 us pulses |
±5%,
0.02 R/min to 600 R/min |
±5%,
2 mR/min to199 R/min |
same |
| Energy
Dependence |
±5%,
30 keV to 1.33 MeV (with build-up material) |
±5%,
10 keV to 40 keV |
(-60)
±5%, 20 keV to 1.33 MeV
(with build-up material)
(-60E)
±5%,
0.2 mm AL HVL to 1.33 MeV (with build-up material) |
| Construction |
Polycarbonate
walls and electrode conductive graphite interior coating;
6 cm3 active volume; 0.05kg |
0.7
mg/cm2 metalized polyester window; polyacetal exterior; 6cm3
active volume; 0.08 kg |
Polycarbonate
walls; conductive graphite exterior coating; 60 cm3 active
volume, 3 m low-noise triax cable; 0.13 kg |
same |
| Chambers |
10x9-3CT |
10x9-180 |
10x9-1800 |
| Min
Rate |
2
µR/s |
.20
nGy/s |
50
nR/s |
500
fGy/s |
5
nR/s. |
50
fGy/s |
| Max
Rate |
40
R/s |
350
mGy/s |
0.6
R/s. |
5.3 mGy/s |
0.02
R/s. |
0.2
mGy/s.. |
| Min
Dose |
20
µR |
200
nGy |
200
nR |
2
nGy |
20
nR . |
200
fGy |
| Max
Dose |
158
kR |
1.4
kGy |
2.6
kR |
23
Gy |
0.26
kR.. |
2.3
Gy |
| Calibration
Accuracy |
±4%
using X-rays @ 150kVp and 10.2
mm AL HVL |
±4%
using X-rays @ 150kVp and 10.2 mm AL HVL |
±4%
using X-rays @ 150kVp and 10.2 mm AL HVL |
| Exposure
Rate Dependence |
±2%,
2mR/s to 40 R/s |
±5%,
20 mR/hr to 2000 R/hr |
+0%,
-5%, 0.1 mR/hr to 20R/hr,
-10% to 65 R/hr |
| Energy
Dependence |
±5%,
3 to 20 mm AL HVL |
±5%,
30 keV to 1.33 MeV (with build up material) |
±5%,
33 keV to 1.33 MeV (with build up material) |
| Construction |
C552
air-equivalent walls and electrode: polyacetal exterior cap;
3 cm3 active volume; 2 m, low noise triax cable; 0.11 kg |
Polycarbonate
walls and electrode; conductive graphite exterior coating;
180 cm3 active volume; 0.11 kg |
Polycarbonate
walls and electrode; conducting graphite exterior coating;
1800 cm3 active volume; 0.54 kg |
| Uniformity
Along Length & Partial Volume Exposure |
±5%,
to within 0.25 cm of chamber ends for a constant volume slice
|
N/A |
N/A |
|
|
|
|
|
|
|
|
9095
SPECIFICATIONS:
Accuracy:
> Dose - ± 4%, ± 1
digit, ± 2 x resolution
> Time - ±0.1%, ±
0.2 msec, ± 1 digit
> Diagnostic kV - ±1kV or
±1%, whichever is greater
> Mammographic kV - ± 0.5
kV
Reproducibility: Dose / Dose Rate: 1% ±
1 digit, ± 2 x resolution
> kV - See kV sensor specifications
Display: 16 character/line, two-line liquid-crystal
Self-Test: (Automatic at turn on) includes display
test, sensor identification, battery check, circuit checks, test
outputs
Setup: Separate setup menu provides user selection
of time and dose units, kV calculation mode, total filtration (affects
kV only), atmospheric corrections, display modes and main-menu items
Controls: Up, Down, Enter, Test, Power
Functions: Dose, Dose Rate, Max Rate, Pulsed, Auto
Dose, Last Dose, kV Pulse, kV Fluoro, kV Dental
Power: 2 C-cells typically provide 40 hours of
operation. Auto power off after 15 minutes of inactivity extends
battery lifetime
Automatic Temperature and Pressure: Included in
overall accuracy specification
Serial Port:
USB
Test: Provides dose and kV simulation
Energy Response: See ion-chamber specifications
Dose Rate: See ion-chamber specifications
kV Filtration dependence: See kV sensor specifications
kV Frequency response:
>
Analog Output: At BNC
-3 dB @ 2.3 kHz
>
kVp Calculation: 0.8
kV up to 4 kHz for 10% ripple, 40-160 kV
Compatibility: Compatible with 10X5 series ion
chambers using adapter 10A95
Compliance: 
EMC: Meets Generic Immunity Standard EN 50082-2
requirement for radiated & conducted susceptibility. Meets EN
55011 requirements for radiated emissions of electromagnetic energy
for Class A devices
Safety: IEC 61010-1
Operating temperature: 15 °C to 35 °C (30
°C for 40x9-MO sensor)
Operating Pressure: 60 to 105 kPa
Humidity: Up to £ 80% RH, (max 20 g/m3),
without condensation
Storage Temperature: -20 °C to +50 °C
Measured Units: /sec, /min, /hr, R, Gy, /cm (for
CT only), kV (peak, practical, average)
Main Cable: Single cable with dual termination.
6 meters
kV
SENSORS
| kV
SENSORS |
40x9-W
Diagnostic kV |
40x9-Mo
Mammography kV |
| Range |
40
- 160 kV |
22
- 40 kV |
| Accuracy |
±
1% or ± 1 kV,
(whichever is greater) |
±
0.5 kV
|
| Filtration
Dependence |
<0.7
kV/mm Al |
22-40
kV <± 0.1 kV/µm, 25 to 35µm Mo 28-40 kV
< 0.8 kV/mm Al |
| Reference
Conditions |
W/Al
(2.5mm), 50 mA, 50 cm SDD |
Mo/Mo
(30 µm), 50 mA, 30 cm SDD |
| Reproductibility |
<0.2
kV COV for 10 measurements |
<
0.1 kV COV for 10 measurements |
| Size
& Weight |
37
mm x 62 mm x 44, 0.4 kg |
37
mm x 62 mm x 44, 0.3 kg |
| 40X9-W
kV Sensor: for Radiographic, Fluoroscopic, CT, Intraoral
and Panoramic Dental kV measurements. |
 |
| 40X9-Mo
kV Sensor: specifically designed for measuring the
kV on mammography (Mo/Mo) machines. |
 |
SOFTWARE
PROGRAM
XLPRO is a Microsoft Excel add-in that provides drop-down menu
control of all 9095 functions and automatic data capture directly
into spreadsheets of your own design. Example templates are also
provided and they are fully customizable to meet your changing
needs. You can store, display and analyze kV, relative dose and
mA waveforms using the tools in the add-in. Fourier transforms
provide accurate ripple measurements; kV waveforms up to 5-sec
may be displayed
XLPRO
Software Waveform Example
(Partial Chart Data
for Graphs Shown) |
SPECIFICATIONS
Full-scale: 2000 mA or 9999 mAs.
mA accuracy: 0.2% of reading at dc plus resolution of +/-0.015 mA
or 3-4 digits.
Absolute value of bi-polar waveform is accurate to 2.5% at 720 Hz.
mAs accuracy (1-s pulse): 5 µAs or 0.2% of reading; resolution
of 0.001 mAs.
Input impedance: Fixed, resistive, 1.7 - 2.7 ohms. 1 ohm sense resistor
plus fuse and wiring.
Input protection: 2-A fuse. Over-voltage protection using 22-J varistor
before the fuse and 17-V Zener diodes after the fuse.
Bandwidth: 2.33 kHz, -3 dB.
ON/OFF: controlled from 9095 control unit. (ON only when a kV function
is active.)
Battery: 9 volt, lifetime is approximately 50 operating hours.
|
 |
 |
|
|
 |