9010/9015
TWO
GREAT METERS,
SAME GREAT PERFORMANCE


| In
their basic form, the 9000 series monitors are multipurpose dosemeters.
Detector choices cover 0.01 µR - 200 kR, 10 µR/h - 1000 kR/h,
over energies of 10 keV - 6 MeV.
Their
traceable accuracy, wide dynamic range, patented technology and
extensive choice of detectors and options make the 9000 series
instruments ideal for countless applications:
- x-ray
Service
- x-ray QA
& Compliance
- x-ray Acceptance
Testing
- Industrial
Applications
- Medical
& Health Physics
- Therapy
QA
- Secondary
Standards
- Nuclear
Applications
With available
options, a 9000 series instrument becomes a multi-purpose tool that
simultaneously determines kVp, time and dose with a single exposure.
Waveform is available for retrieval. |
System
2000D Series
( 9010 with 4083 Accu-kV)

(micro-laptop not included)
Typical
medical physics applications include:
-
Radiography
-
Fluoroscopy
-
Angiography,
Mammography
-
Image
Intensifier
-
Pulsed
x-ray
-
Cine
-
CT
-
Dental
Other
applications include:
-
137Cs
-
60Co
Irradiator measurements
-
Survey
meter calibration
-
Industrial
x-ray (cables up to three hundred feet)
-
Very
low dose rates and very high dose rate
applications (up to 107R/hr with special modification)
|
Automatic corrections
and clearly labeled controls are all designed to simplify operation
and minimize errors. The end result is a measurement of undisputed
accuracy which requires no corrections.
9010
Compatibility with all existing 10X series chambers is insured via the
9060 converter (order the 9010RM-S monitor which includes the 9060).
The rugged 20V30 case can house the 9010, three Radcal chambers, a kVp
detector, cables, the IR remote, test stand, manual and an HVL kit.
Built with the Radcal model 2025 user in mind, the 9010 accepts inputs
from all existing 20X5 series sensors via the 90A25 cable adapter (order
the 9010M-S along with the 90A25).
9015

Much
like the original 'MDH' (the 1015C), the 9015 is self-contained in
a rugged metal case. It can use all existing 10X series chambers.
Although it is only about half the size of the 1015, the 9015 can
house up to two chambers, the 9060 converter and a rugged 6 meter
cable with metal connectors. All of this with Radcal's ionization
chamber technology - The Gold Standard.
9010
vs. 9015
Although differing
in appearance, the 9010 and 9015 are electronically identical. The difference
is only in their exterior housing.
FUNCTIONS
Auto-Zero:
When on, this mode provides the user with improved low-level measurements
by automatically subtracting background signals.
Dose
rate: Threshold-free rate mode. Update period is adjusted automatically
for best accuracy.
Max
Dose rate: Works like the peak hold function of a digital volt meter.
Perfect for setting the maximum rate on a fluoro x-ray machine.
Dose:
Threshold-free integrated dose mode. Displays the integrated dose. Updates
once per second.
Auto
Dose: Threshold triggered exposure mode. Provides hands off operation
for reproducibility measurements.
Last
Dose: Integrates total dose continuously.
Pulsed
(Cine): Displays the reading directly in exposure/frame (0.25-90
f/sec). User selectable delays and averaging periods.
SELF
TESTING
Just
connect a compatible sensor and turn on. A comprehensive startup test
verifies full functionality of the instrument.
- LCD Display
Test LCD display verification.
- Battery Test
Checks battery voltage
- Chamber Voltage
Test Checks bias voltage
DIRECT
READING
- Auto-chamber
recognition: Applies chamber corrections
- Auto-temperature
compensation: Corrects for temperature
- Auto-pressure
compensation: Corrects for pressure
- Minimal energy
corrections: Regardless of filtration or target material
- Dedicated
Pulsed(Cine) mode: Reads directly in dose/frame
- ...In the units
of your choice: R, Gy, C/kg, Sv, /sec, /min, /hr
MULTIFUNCTION
- kVp, time &
dose capture for a variety of modalities
- Continuous
kVp & dose rate capture for continuous measurements
- The 9010 or
9015 become multi-function when used in conjunction with a Radcal
Non-
Invasive kVp and Time instrument
AUTOMATED
- RS232 Serial
Port The bi-directional serial port permits hands-off operation
from a remote computer. Excel_ add-in software is available
SPECIFICATIONS
| 9010
& 9015 |
| Accuracy: |
±4%
of reading, ±1 digit, ±2 times resolution |
| Repeatability: |
±1%
of reading, ±1 digit, ±2 times resolution |
| Setup: |
Easy
to access setup menu allows choice of main-menu functions, selection
of radiation units (R, Gy, Sv, C/kg), time units (/sec, /min, /hr),
frame (pulse) rates, auto-power off delay (0-99 minutes or disabled).
Temperature & Pressure information. Additional set-up items
automatically become available when options are installed. |
| Functions: |
Dose,
Dose Rate, Max. Rate, Pulsed, Auto Exposure, Last Exposure, Auto-zero,
kVp Pulse (opt.), kVp Fluoro (opt.) |
| Auto
Temperature: |
Accuracy
2% |
| Compatibility |
Fully
compatible with 10X series chambers. 20X series sensors do not support
Pulsed, Auto-Dose & Auto-Zero modes |
| Auto
Pressure: |
Accuracy
0.2%,+2% of reference pressure - ambient (optional, specify Opt-P) |
| Analog
Capture: |
0-5
V, 10 kHz, 12-bit analog to digital signal capture. 100 µsec
and 1.25 mV resolution over 1 sec. Playback to analog output. Data
available at the RS-232 port or ANALOG OUT. Timing 5 ms to 6.5 sec
with 100 µsec resolution with ±0.05% accuracy exclusive
of non-invasive detector (optional, specify Opt-C). kV specifications
depend on kV sensor. |
| Infrared
remote: |
Allows
full control from up to 10 m away. Will go through lead glass. (optional,
specify Opt-R) |
Optional
Features
Option-C
(Analog Signal Capture)
This option, when used in conjunction with a suitable non-invasive kV
sensor and an ionization chamber, turns the 9000 series into multi-function
meters that determine kVp, time and dose with a single exposure. The
kV waveform is also recorded and may be displayed on an oscilloscope
or downloaded to a computer and viewed using 9000 Excel Software Program.
Note:
The cost of the kV sensor is exclusive of the cost of the option C.
You may purchase a compatible sensor from Radcal; contact us for details.
kVp
Pulse: To be used with exposures
lasting <6 s such as those encountered in mammography and radiography.
kVp
Fluoro: Provides continuous display of the kVp and dose rate
with updates occurring at 1-4 sec intervals.
Option-P
(Auto-Pressure Compensation)
The barometric pressure compensation circuit provides automatic correction
of all ion chamber readings and displays the measured pressure. Perfect
for mountainous regions. The automatic pressure compensation may be
overridden. Please note, automatic temperature compensation is a standard
9000 series feature.
XLPRO-9040 SOFTWARE PROGRAM
XLPRO-9040 is a Microsoft Excel add-in that provides drop-down menu
control of all 9000 series functions and automatic data capture directly
into spreadsheets of your own design. Example templates are also provided
and they are fully customizable to meet your changing needs. Computer
requirements: spare serial port and Microsoft Windows 9x, NT 4.x, 2000,
ME or XP and Microsoft Excel 97 or later.
IONIZATION
CHAMBERS
All
Radcal ion chambers are designed to provide the user with superior energy
response characteristics, unparalleled in the industry. The resulting
chamber accuracy and consistent reliability are well documented in the
technical literature1-5.
 |
The
Commission of the European Communities undertook what proved to
be the largest, most comprehensive intercomparison of diagnostic
dose-meters ever done. Our meters, because of their ease of use
and linear energy response, were found to provide the users with
consistently good results. Please ask us for your free copy. |
CHAMBER
SELECTION
CONNECTOR
Contains the chamber factory calibration as well as the chamber volume
information for the detector. This provides chamber interchangeability
between instruments without re-calibration.
90X
FAMILY OF CHAMBERS
Ideal for single detector or existing 90X series based systems. Please
contact us.

Click numbers for enlarged product views |
CHAMBER
Click
on each item listed above and it will link to the proper specifications.
|
10.
DIN 23.04/95.09 The 9060 CONVERTER
Radcal radiation monitors use a patented current -to-frequency converter
that digitizes the ion chamber output. This provides a wide dynamic
range and minimizes noise and stability problems.
CHAMBER
OVERVIEW - ROENTGENS
| Sensor |
Rate
Resolution |
Maximum |
Dose
Resolution |
Maximum |
| 10X5-0.18 |
0.1 R/min |
650 kR/hr |
0.01 mR |
2.3 MR |
| 10X5-0.6 |
1 mR/min |
8 kR/min |
0.01 mR |
700 kR |
| 10X5-6 |
0.1 mR/min |
840 R/min |
1 µR |
70 kR |
| 10X5-3CT |
0.1 mR/min
|
1.7 kR/min |
1 µR |
140 kR |
| 10X5-6M |
0.1 mR/min |
840 R/min |
1 µR |
70 kR |
| 10X5-60,60E |
0.01 mR/min |
80 R/min |
0.1 µR |
7 kR |
| 10X5-180 |
0.1 mR/hr |
1.7 kR/hr |
0.01 µR |
2.3 kR |
| 10X5-1800
|
0.01 mR/hr |
65 R/hr |
0.1 µR |
230 R |
CHAMBER
OVERVIEW - GRAYS
| Sensor |
Rate
Resolution |
Maximum |
Dose
Resolution |
Maximum |
| 10X5-0.18 |
1
mGy/hr |
5.7
kGy/hr |
0.1
µGy |
20
kGy |
| 10X5-0.6 |
1
µGy/min |
74
Gy/min |
0.1
µGy |
6
kGy |
| 10X5-6 |
0.1
µGy/min |
7.4
Gy/min |
0.01
µGy |
600
Gy |
| 10X5-3CT |
1
µGy/min |
15
Gy/min |
0.01
µGy |
1.2
kGy |
| 10X5-6M |
0.1
µGy/min |
7.4
Gy/min |
0.01
µGy |
600
Gy |
| 10X5-60,60E |
0.01
µGy/min |
730
mGy/min |
1
nGy |
60
Gy |
| 10X5-180 |
1
µGy/hr |
15
Gy/hr |
0.1
nGy |
20
Gy |
| 10X5-1800 |
0.1
µGy/hr |
575
mGy/hr |
0.01
nGy |
2
Gy |
10X5-60
&10X5-60E
SERVICE CHAMBERS
The dynamic
range and thin profile of these chambers make them ideal for
virtually all X-ray service applications
-
Input
Dose at the Image Intensifier
-
High
dose rates encountered in Fluoroscopy
-
Cine,
spot film devices & other special procedures
Additionally,
the -60E(xtended) chamber's increased sensitivity at lower energies
turns the chamber into a "Universal" detector, covering mammography
through R&F and beyond.
The 8231
holder is recommended.
| |
| Rate
Specifications: |
0.1
mR/min - 80 R/min
0.01µGy/min - 730 mGy/min |
| Exposure
Specifications: |
0.1µR
- 7 kR
1 nGy - 60 Gy |
| Cine
Specifications: |
0.01
µR/f - >100 mR/f
0.1 nGy/f - >1.0 mGy/f |
| Calibration
Accuracy (-60): |
±4%
using X-rays @ 150 kVp & 10.2 mm Al HVL |
| Calibration
Accuracy (-60E): |
±4%
using X-rays @ 50 kVp & 0.88 mm Al HVL |
| Exposure
Rate Dependence: |
±5%,
2 mR/min to 199 R/min |
| Energy
Dependence (-60): |
±5%,
20 keV to 1.33 MeV(with build-up material) |
| Energy
Dependence (-60E): |
±5%,
0 .2mm AL HVL to 1.33 MeV (with build-up material) |
| Construction:
|
Polycarbonate
walls; conductive graphite exterior coating; 60 cm3 active volume,
3 m low-noise triax cable; 0.13 kg |
10X5-6
GENERAL PURPOSE, IN BEAM CHAMBER
The 10X5-6 chamber
permits in-beam measurements as described in "Routine Compliance Testing
for Diagnostic x-ray systems" (PB89-205215). This chamber is ideal
when using the CDRH geometry (the 10T5 test stand).
SPECIFICATIONS
| |
| Rate
Specifications |
0.1 mR/min - 840 R/min
0.1 µGy/min - 7.4 Gy/min |
| Exposure
Specifications |
1 µR - 70 kR
0.1µGy - 600 Gy |
| Cine
Specifications: |
0.1
µR/f - >1R/f
1 nGy/fr - >10 mGy/f |
| Calibration
Accuracy: |
±4%
using X-rays @ 60 kVp & 2.8 mm Al HVL |
| Exposure
Rate Dependence: |
±5%,
0.02 to 1000 R/min, up to 500R/s for 10 ms pulses |
| Energy
Dependence |
±5%,
30 keV to 1.33 MeV (with build-up material) |
| Construction: |
Polycarbonate
walls and electrode conductive graphite interior coating;
6 cm3 active volume; 0.05 kg |
10X5-6M
DEDICATED MAMMOGRAPHY CHAMBER
Long
before the Mammography Quality Assurance Standards Act, Radcal designed
what proved to be one of the finest low energy chambers on the market.
It's extraordinary flat energy response over the 10 - 40 keV has been
documented in technical papers and makes corrections unnecessary.
Unlike solid state detectors, the -6M's response is not influenced
by tube target material or filtration.
SPECIFICATIONS
| |
| Rate
Specifications: |
0.1
mR/min - 840 kR/min
0.1 µGy/min - 7.4 Gy/min |
| Exposure
Specifications: |
1
µR - 70 kR
0.01 µGy - 600 Gy |
| Calibration
Accuracy: |
±4%
using X-rays @ 20 kVp & 0.26 mm Al HVL |
| Exposure
Rate Dependence |
±5%,
0.02 R/min to 600 R/min |
| Energy
Dependence: |
±5%,
10 keV to 40 keV |
| Construction: |
0.7
mg/cm2 metalized polyester window; polyacetal exterior; 6 cm3
active volume; 0.08 kg |
10X5-180
LEAKAGE AND LOW LEVEL MEASUREMENTS CHAMBER
The
10X5-180 ion chamber is designed for leakage measurements as described
in "Regulations for the Administration and Enforcement of The Radiation
Control for Health and Safety Act of 1968" (HHS Publication FDA 88-8035),
and provides the required effective cross-section of 100 cm2.
SPECIFICATIONS
| |
| Rate
Specifications: |
0.1 mR/hr - 1.7 R/hr
1 µGy/hr - 15 Gy/hr |
| Exposure
Specifications: |
0.01 µR - 2.3 kR
0.1 nGy - 20 Gy |
| Calibration
Accuracy: |
±4%
using X-rays @ 150 kVp & 10.2 mm Al HVL |
| Exposure
Rate Dependence |
±5%, 20 mR/hr to 2000 R/hr |
| Energy
Dependence: |
±5%,
33 keV to 1.33 MeV (with build-up material) |
| Construction: |
Polycarbonate walls and electrode; conducting graphite exterior
coating; 180 cm3 active volume; 0.11 kg |
10X5-1800
RADIATION PROTECTION CHAMBER
The
10X5-1800 ionization chamber is intended for low-level radiation measurements
such as shielding leakage, irradiator and environmental. Unlike typical
survey meters, the 10X5-1800 chamber offers improved accuracy over a
wider dynamic range.
SPECIFICATIONS
| |
| Rate
Specifications: |
0.01
mR/hr - 65 R/hr
0.1 µGy/hr - 575 mGy/hr |
| Exposure
Specifications: |
1
nR - 230 R
0.01 nGy - 2 Gy |
| Calibration
Accuracy: |
±4%
using X-rays @ 150 kVp & 10.2 mm Al HVL |
| Exposure
Rate Dependence |
+0%,
-5%, 0.1 mR/hr to 20 R/hr, -10% to 65 R/hr |
| Energy
Dependence: |
±5%,
33 keV to 1.33 MeV (with build-up material) |
| Construction: |
Polycarbonate
walls and electrode; conducting graphite exterior coating; 1800
cm3 active volume; 0.54 kg |
(CTDI)
10X5-3CT
CHAMBER FOR COMPUTED TOMOGRAPHY DOSE INDEX
When
used in conjunction with Radcal's 20CT6 (head phantom) and Radcal's
20CT14 (body phantom) the 10x5-3CT permits evaluation of the radiation
output as prescribed by HHS publication FDA 88-8035. It's excellent
energy, partial volume response and position uniformity has been well
documented 4,5.
SPECIFICATIONS
| |
| Rate
Specifications: |
0.1
mR/min - 1.7 kR/min
1 µGy/min - 15 Gy/min |
| Exposure
Specifications: |
1µR
- 140 kR
0.01 µGy - 1.2 kGy |
| Calibration
Accuracy: |
±4%
using X-rays @ 150 kVp and 10.2 mm Al HVL |
| Exposure
Rate Dependence |
±2%,
2 mR/s to 40 R/s |
| Energy
Dependence: |
±5%,
3 - 20 mm Al HV |
| Uniformity
Along Length & Partial Volume Response |
±5%,
to within 0.25 cm of chamber ends for a constant volume slice |
| Construction: |
C552
air-equivalent walls and electrode; polyacetal exterior cap; 3 cm3
active volume; 2 m, low-noise triax cable; 0.11 kg |
1OX5-0.6
THERAPY QA CHAMBER
This
high dose rate chamber provides an excellent response at therapy and
other high energy, high dose rate applications. The fully guarded chamber
is mounted at the end of a 12 m low noise triax cable.
SPECIFICATIONS
| |
| Rate
Specifications: |
1
mR/min - 8 kR/min
1 µGy/min - 74 Gy/min |
| Exposure
Specifications: |
0.01
mR - 700 MR
0.1 µGy - 6 kGy |
| Calibration
Accuracy: |
±4%
using 60Co |
| Exposure
Rate Dependence |
±2%,
0.6 R/min to 6 kR/min |
| Energy
Dependence: |
±5%,
40 keV to 1.33 MeV (with build-up cap) |
| Construction: |
C552
air-equivalent material & electrode; polyacetal exterior cap; 0.6
cm3 active volume; 12 m, low-noise triax cable; 0.28 kg |
10X5-0.18
HIGH DOSE-RATE CHAMBER
The
10X5-0.18 ion chamber is intended for in-beam measurements of high-intensity
gamma radiation. It is suitable for cavity gamma irradiators as well
as beam type irradiators. The fully guarded chamber is mounted at the
end of a 3 meter, low-noise triax cable.
SPECIFICATIONS
| |
| Rate
Specifications: |
0.1
R/hr - 650 kR/hr
1 mGy/hr - 5.7 kGy/hr |
| Exposure
Specifications: |
0.01
mR - 2.3 MR
0.1 µGy - 20 kGy |
| Calibration
Accuracy: |
±4%
using 60Co |
| Exposure
Rate Dependence |
±2%,
10 R/hr to 650 kR/hr |
| Energy
Dependence: |
±5%,
45 keV to 1.33 MeV |
| Construction: |
C552
air-equivalent walls & electrode; polyacetal exterior cap; 0.18
cm3 active volume; 3 m, low-noise triax cable; 0.1 kg |
Radiation Protection
Dosimetry, Volume 43, No. 1/4, Studies On the Performance of Diagnostic
Air Kerma Meters in the US, L.K. Wagner
Radiation Protection
Dosimetry, Volume 43, No. 1/4, Diagnostic Dosimeter Intercomparison,
Commission of European Communities
Medical Physics,
17(6), Nov/Dec 90, On the measurement of HVL in Film Screen Mammography,
L. Wagner, F. Cerra, B. Archer
Health Physics
Journal, Vol 37 (Dec), pp. 773-774, Evaluation of the MDH Industries,
Inc, Pencil Chamber for Direct Beam CT Measurements, William Pavlicek,
John Horton, Robert Turco
Measurement
of Radiation for X-Ray Computed Tomoghraphy Using Pencil-Shaped Ionization
Chambers, R.M. Gagne, T.B. Shope, J.L. McCrohan and C.K. Showalter