4083 Non-Invasive Analyzer
From the
manufacturer of the Dynalyzer, the standard in invasive kV
measurements, comes a new generation of instruments using Accu-kV
Technology. |
- Large
Easy-to-Read Display
- Versatility
...Stand
alone, Connect to Scope,Computer or Radcal 9000 dose meter
- Extremely
Accurate
...kVp-Time-Pulses
- Compact
Sensor for less Interference while AEC is in use
- Advanced
kVp Algorithms
...Accu-"Peak"
Ripple Corrected kVp
...Peak
Averaging
...FFT
Algorithms accurately distinguish true waveform ripple from noise
|
Feature
Rich Performance:
- Accu-kV
Technology
- Wide range
with a single sensor
- Extremely
accurate in presence of low mA or noisy signals
- No linearity
corrections required
- Fully automatic
range changing - no need for multiple filter packs or manual filter
changes
- No need for
complicated k-Edge measurements to achieve ±0.5 accuracy
in the mammo energy range.
- Interchangeable
kV sensors with any 4083 Analyzer, 4082 or 4081 Control Unit.
- User adjustable
threshold to avoid pre-heat pulse common to dental x-ray machines.
|
4083
kV Analyzer
*computer
not included; optional
4083 software sold separately |
Computer
Output Example
*XLPRO
Microsoft Excel Add-In |
 |
|
| |
|
XLPRO
Software Template Example |
Partial
Data for Adjacent Graph |
 |
|
40X5-W
Diagnostic kV Sensor
For
Radiographic, Fluoroscopic, CT, Interoral & Panoramic Dental kV Measurements

40X5-W
Diagnostic kV Sensor |
| Specifications |
|
40
- 160 kV |
| Range: |
|
|
| Accuracy: |
±
1% or ± 1 kV, whichever is greater,
<0.7 kV/mm Al |
|
2.89 mmAL HVL @ 70 kVp , 50 mA, 50 cm SDD |
| Reference
Conditions: |
|
<0.2
kV COV of 10 measurements |
| Reproducibility: |
|
7
mm x 62 mm x 44, o.4 kg |
| Size
& Weight: |
40X5-Mo
Mammography kV Sensor |
|
| Specifications |
|
| Range: |
22
- 40 kV |
|
| Accuracy: |
±
0.5 kV
22-40 kV <± 0.1 kV/µm, 25 to 35µm Mo
28-40 kV < 0.8 kV/mm Al |
|
| Reference
Conditions: |
Mo/Mo
(30 µm), 50 mA, 30 cm SDD |
|
| Reproducibility: |
<
0.1 kV COV of 10 measurements |
|
| Size
& Weight: |
37
mm x 62 mm x 44, 0.3 kg |
|
4083
Accu kV Non-Invasive Analyzer |
| Specifications |
|
0-5
V, 30.52 mV / kV |
| Output
(BNC): |
|
|
| Output
(USB): |
Optional
XLPRO-USB8 Microsoft Excel Add-In required, RS-232 to USB Adapter
included |
|
Four
push buttons (Off, Normal, Fluoro, Dental) |
| Controls: |
|
|
| Indicators: |
Green
(READY), LEDs Four-digit numeric display w/units and mode indications |
|
CE |
| Compliance: |
|
|
|
6
externally accessible AA batteries (1.5V, IEC#LR6), ~40 hrs |
| Power: |
|
|
|
15°
to 30° C <90% RH without condensation |
| Operating: |
|
|
|
-35°
to +50° C |
| Storage: |
|
|
|
130
x 210 x 75 mm, 1.1 kg |
| Size
& Weight: |
|
|