DISCONTINUED
MODEL
Specifications for reference only

For
Radiology Quality Assurance
The
3036 Advantages ...
- Cost effective,
easy to use compact unit.
- Use for both above
and below table X-ray sources with the built-in bi-directional
ion chamber.
- Hands off
operation in Pulse mode for dose and time in a single exposure.
- Rate mode is available
for fluoroscopic diagnostic measurements.
- Battery operated
and automatic shut-down after ten minutes of inactivity.
- A built-in test
function permits verification of operation for all functions.
- Can be used for
X-ray machines with AEC by employing optional external ion chamber.
SPECIFICATIONS
| 3036 |
| Range
- Rate Mode: |
0.001
to 1200 R/min (0.01 to 999.9 mGy/min) |
| Pulse
Mode: |
0.0001
to 2.50 R (0.0001 to 22.00 mGy)
1 to 9999 ms. Resolution 0.1 sec. Pulse
exposure threshold occurs at 600 mR/min
and ends at 50% of the maximum rate. Automatic reset displays the
last measurement until a new pulse exceeds 600 mR/min. |
| Exposure
Mode: |
0.0001
to 670 R (0.0001 to 5900 mGy) |
| Accuracy: |
±5%
and ±1 ms (4 ms @ 600 mR/min) |
| Maximum
Exposure Rate: |
1200
R/min |
| Energy
Response: |
±5%,
1.5 to 14mm Al hvl |
| Detector
- Measurement Area: |
13
cm2 (41mm diameter) |
| Center
of Ion Chamber: |
38
mm from top |
| Bias
Battery: |
67.5
V (>1 yr life) |
3036
Dosimeter with Options

30X6-6M External Chamber (Right)
30A6-Chamber Adapter with 10X5 series Chamber (Left)
| 10X5-series
Ion chambers |
| 10X5-6 |
Service |
| 10X5-6M |
Mammography
|
| 10X5-60 |
General
Purpose, In Beam |
| 10X5-180 |
Leakage
and low level |
| 10X5-0.6 |
Therapy
QA |
| 10X5-3CT |
Computed
Tomography Dose Index (CTDI) |
See individual chamber
specifications below photos
Model
3036 Options
30X6-11
External Chamber
 |
30X6-6M
External Chamber
 |
General
R&F, in-beam, 11 cm3 active volume with 1 m low-noise triax.
|
Mammography,
6 cm3 active volume with 3 m low-noise triax. Maximum Dose Rate
is 100 R/min.
|
| Both
the 30X6-11 External Chamber and 30X6-6M External Chamber connect
to back panel of unit. All other 3036 Specifications apply. |
|
30A6-Chamber
Adapter

Chamber
Adapter
Size: 6 x 6 x 12 cm
Weight: 200 gm
Cable: 1 meter
The 30A6 Multi-Chamber Adapter permits the 3036 to be used in
a wide variety of applications. Six different 10X5-series ion
chambers can be used with the adapter. Set the selector switch
to the chamber being used and the 3036 provides direct readout.
For the 10X5-60 and 10X5-180 ion chambers only, the R is read
as mR. Measurement accuracy with the Adapter is ± 10%.
The 30A6 Adapter is not designed to provide accurate exposure
time readings in Pulse mode with the 10X5 series chambers.
|
| Chamber
Specifications*
Measurement Range for 10x5-Series Ion Chambers |
| 1 |
6 |
6M |
60 |
180 |
0.6 |
3CT |
| Rate: |
0.001
R/min
1500 R/min |
0.001
R/min
100 R/min |
0.1
mR/min
9999 mR/min |
0.1
mR/min
9999 mR/min |
0.01
R/min
9999 R/min |
0.01
R/min
4100 R/min |
| Exposure: |
0.0001
R
1160 R |
0.0001
R
1160 R |
0.01
mR
9999 mR |
0.001
mR
9999 mR |
0.001
R
9999 R |
0.0001
R
2300 R |
* Note: the
mode 30A6 Adapter is not designed to provide accurate exposure time readlings,
in Pulse mode, with the 10X5 series chambers.
|
|
 |