CHAMBER
SELECTION
CONNECTOR
Contains the chamber factory calibration as well as the chamber volume
information for the detector. This provides chamber interchangeability
between instruments without re-calibration.
90X
FAMILY OF CHAMBERS
Ideal for single detector or existing 90X series based systems. Please
contact us.
Click numbers for enlarged product views |
CHAMBER
Click
on each item listed above and it will link to the proper specifications.
|
10.
DIN 23.04/95.09 The 9060 CONVERTER
Radcal radiation monitors use a patented current -to-frequency converter
that digitizes the ion chamber output. This provides a wide dynamic range
and minimizes noise and stability problems.
CHAMBER
OVERVIEW - ROENTGENS
| Sensor |
Rate
Resolution |
Maximum |
Dose
Resolution |
Maximum |
| 10X5-0.18 |
0.1 R/min |
650 kR/hr |
0.01 mR |
2.3 MR |
| 10X5-0.6 |
1 mR/min |
8 kR/min |
0.01 mR |
700 kR |
| 10X5-6 |
0.1 mR/min |
840 R/min |
1 µR |
70 kR |
| 10X5-3CT |
0.1 mR/min
|
1.7 kR/min |
1 µR |
140 kR |
| 10X5-6M |
0.1 mR/min |
840 R/min |
1 µR |
70 kR |
| 10X5-60,60E |
0.01 mR/min |
80 R/min |
0.1 µR |
7 kR |
| 10X5-180 |
0.1 mR/hr |
1.7 kR/hr |
0.01 µR |
2.3 kR |
| 10X5-1800
|
0.01 mR/hr |
65 R/hr |
0.1 µR |
230 R |
CHAMBER
OVERVIEW - GRAYS
| Sensor |
Rate
Resolution |
Maximum |
Dose
Resolution |
Maximum |
| 10X5-0.18 |
1
mGy/hr |
5.7
kGy/hr |
0.1
µGy |
20
kGy |
| 10X5-0.6 |
1
µGy/min |
74
Gy/min |
0.1
µGy |
6
kGy |
| 10X5-6 |
0.1
µGy/min |
7.4
Gy/min |
0.01
µGy |
600
Gy |
| 10X5-3CT |
1
µGy/min |
15
Gy/min |
0.01
µGy |
1.2
kGy |
| 10X5-6M |
0.1
µGy/min |
7.4
Gy/min |
0.01
µGy |
600
Gy |
| 10X5-60,60E |
0.01
µGy/min |
730
mGy/min |
1
nGy |
60
Gy |
| 10X5-180 |
1
µGy/hr |
15
Gy/hr |
0.1
nGy |
20
Gy |
| 10X5-1800 |
0.1
µGy/hr |
575
mGy/hr |
0.01
nGy |
2
Gy |
10X5-60
&10X5-60E
SERVICE CHAMBERS
The dynamic
range and thin profile of these chambers make them ideal for virtually
all X-ray service applications
-
Input
Dose at the Image Intensifier
-
High
dose rates encountered in Fluoroscopy
-
Cine,
spot film devices & other special procedures
Additionally,
the -60E(xtended) chamber's increased sensitivity at lower energies
turns the chamber into a "Universal" detector, covering mammography
through R&F and beyond.
The 8231 holder
is recommended.
| |
| Rate
Specifications: |
0.1
mR/min - 80 R/min
0.01µGy/min - 730 mGy/min |
| Exposure
Specifications: |
0.1µR
- 7 kR
1 nGy - 60 Gy |
| Cine
Specifications: |
0.01
µR/f - >100 mR/f
0.1 nGy/f - >1.0 mGy/f |
| Calibration
Accuracy (-60): |
±4%
using X-rays @ 150 kVp & 10.2 mm Al HVL |
| Calibration
Accuracy (-60E): |
±4%
using X-rays @ 50 kVp & 0.88 mm Al HVL |
| Exposure
Rate Dependence: |
±5%,
2 mR/min to 199 R/min |
| Energy
Dependence (-60): |
±5%,
20 keV to 1.33 MeV(with build-up material) |
| Energy
Dependence (-60E): |
±5%,
0 .2mm AL HVL to 1.33 MeV (with build-up material) |
| Construction:
|
Polycarbonate
walls; conductive graphite exterior coating; 60 cm3 active volume,
3 m low-noise triax cable; 0.13 kg |
10X5-6
GENERAL PURPOSE, IN BEAM CHAMBER
The 10X5-6 chamber
permits in-beam measurements as described in "Routine Compliance Testing
for Diagnostic x-ray systems" (PB89-205215). This chamber is ideal when
using the CDRH geometry (the 10T5 test stand).
SPECIFICATIONS
| |
| Rate
Specifications |
0.1 mR/min - 840 R/min
0.1 µGy/min - 7.4 Gy/min |
| Exposure
Specifications |
1 µR - 70 kR
0.1µGy - 600 Gy |
| Cine
Specifications: |
0.1
µR/f - >1R/f
1 nGy/fr - >10 mGy/f |
| Calibration
Accuracy: |
±4%
using X-rays @ 60 kVp & 2.8 mm Al HVL |
| Exposure
Rate Dependence: |
±5%,
0.02 to 1000 R/min, up to 500R/s for 10 ms pulses |
| Energy
Dependence |
±5%,
30 keV to 1.33 MeV (with build-up material) |
| Construction: |
Polycarbonate
walls and electrode conductive graphite interior coating;
6 cm3 active volume; 0.05 kg |
10X5-6M
DEDICATED MAMMOGRAPHY CHAMBER
Long
before the Mammography Quality Assurance Standards Act, Radcal designed
what proved to be one of the finest low energy chambers on the market.
It's extraordinary flat energy response over the 10 - 40 keV has been
documented in technical papers and makes corrections unnecessary. Unlike
solid state detectors, the -6M's response is not influenced by tube
target material or filtration.
SPECIFICATIONS
| |
| Rate
Specifications: |
0.1
mR/min - 840 kR/min
0.1 µGy/min - 7.4 Gy/min |
| Exposure
Specifications: |
1
µR - 70 kR
0.01 µGy - 600 Gy |
| Calibration
Accuracy: |
±4%
using X-rays @ 20 kVp & 0.26 mm Al HVL |
| Exposure
Rate Dependence |
±5%,
0.02 R/min to 600 R/min |
| Energy
Dependence: |
±5%,
10 keV to 40 keV |
| Construction: |
0.7
mg/cm2 metalized polyester window; polyacetal exterior; 6 cm3 active
volume; 0.08 kg |
10X5-180
LEAKAGE AND LOW LEVEL MEASUREMENTS CHAMBER
The 10X5-180
ion chamber is designed for leakage measurements as described in "Regulations
for the Administration and Enforcement of The Radiation Control for Health
and Safety Act of 1968" (HHS Publication FDA 88-8035), and provides the
required effective cross-section of 100 cm2.
SPECIFICATIONS
| |
| Rate
Specifications: |
0.1 mR/hr - 1.7 R/hr
1 µGy/hr - 15 Gy/hr |
| Exposure
Specifications: |
0.01 µR - 2.3 kR
0.1 nGy - 20 Gy |
| Calibration
Accuracy: |
±4%
using X-rays @ 150 kVp & 10.2 mm Al HVL |
| Exposure
Rate Dependence |
±5%, 20 mR/hr to 2000 R/hr |
| Energy
Dependence: |
±5%,
33 keV to 1.33 MeV (with build-up material) |
| Construction: |
Polycarbonate walls and electrode; conducting graphite exterior coating;
180 cm3 active volume; 0.11 kg |
10X5-1800
RADIATION PROTECTION CHAMBER
The 10X5-1800
ionization chamber is intended for low-level radiation measurements such
as shielding leakage, irradiator and environmental. Unlike typical survey
meters, the 10X5-1800 chamber offers improved accuracy over a wider dynamic
range.
SPECIFICATIONS
| |
| Rate
Specifications: |
0.1
mR/hr - 65 R/hr
0.1 µGy/hr - 575 mGy/hr |
| Exposure
Specifications: |
1
nR - 230 R
0.01 nGy - 2 Gy |
| Calibration
Accuracy: |
±4%
using X-rays @ 150 kVp & 10.2 mm Al HVL |
| Exposure
Rate Dependence |
+0%,
-5%, 0.1 mR/hr to 20 R/hr, -10% to 65 R/hr |
| Energy
Dependence: |
±5%,
33 keV to 1.33 MeV |
| Construction: |
Polycarbonate
walls and electrode; conducting graphite exterior coating; 1800 cm3
active volume; 0.54 kg |
(CTDI)
10X5-3CT
CHAMBER FOR COMPUTED TOMOGRAPHY DOSE INDEX
When
used in conjunction with Radcal's 20CT6 (head phantom) and Radcal's 20CT14
(body phantom) the 10x5-3CT permits evaluation of the radiation output
as prescribed by HHS publication FDA 88-8035. It's excellent energy, partial
volume response and position uniformity has been well documented 4,5.
SPECIFICATIONS
| |
| Rate
Specifications: |
0.1
mR/min - 1.7 kR/min
1 µGy/min - 15 Gy/min |
| Exposure
Specifications: |
1µR
- 140 kR
0.01 µGy - 1.2 kGy |
| Calibration
Accuracy: |
±4%
using X-rays @ 150 kVp and 10.2 mm Al HVL |
| Exposure
Rate Dependence |
±2%,
2 mR/s to 40 R/s |
| Energy
Dependence: |
±5%,
3 - 20 mm Al HV |
| Uniformity
Along Length & Partial Volume Response |
±5%,
to within 0.25 cm of chamber ends for a constant volume slice |
| Construction: |
C552
air-equivalent walls and electrode; polyacetal exterior cap; 3 cm3
active volume; 2 m, low-noise triax cable; 0.11 kg |
1OX5-0.6
THERAPY QA CHAMBER
This
high dose rate chamber provides an excellent response at therapy and other
high energy, high dose rate applications. The fully guarded chamber is
mounted at the end of a 12 m low noise triax cable.
SPECIFICATIONS
| |
| Rate
Specifications: |
1
mR/min - 8 kR/min
1 µGy/min - 74 Gy/min |
| Exposure
Specifications: |
0.01
mR - 700 MR
0.1 µGy - 6 kGy |
| Calibration
Accuracy: |
±4%
using 60Co |
| Exposure
Rate Dependence |
±2%,
0.6 R/min to 6 kR/min |
| Energy
Dependence: |
±5%,
40 keV to 1.33 MeV (with build-up cap) |
| Construction: |
C552
air-equivalent material & electrode; polyacetal exterior cap; 0.6
cm3 active volume; 12 m, low-noise triax cable; 0.28 kg |
10X5-0.18
HIGH DOSE-RATE CHAMBER
The 10X5-0.18
ion chamber is intended for in-beam measurements of high-intensity gamma
radiation. It is suitable for cavity gamma irradiators as well as beam
type irradiators. The fully guarded chamber is mounted at the end of a
3 meter, low-noise triax cable.
SPECIFICATIONS
| |
| Rate
Specifications: |
0.1
R/hr - 650 kR/hr
1 mGy/hr - 5.7 kGy/hr |
| Exposure
Specifications: |
0.01
mR - 2.3 MR
0.1 µGy - 20 kGy |
| Calibration
Accuracy: |
±4%
using 60Co |
| Exposure
Rate Dependence |
±2%,
10 R/hr to 650 kR/hr |
| Energy
Dependence: |
±5%,
45 keV to 1.33 MeV |
| Construction: |
C552
air-equivalent walls & electrode; polyacetal exterior cap; 0.18 cm3
active volume; 3 m, low-noise triax cable; 0.1 kg |
Radiation Protection
Dosimetry, Volume 43, No. 1/4, Studies On the Performance of Diagnostic
Air Kerma Meters in the US, L.K. Wagner
Radiation Protection
Dosimetry, Volume 43, No. 1/4, Diagnostic Dosimeter Intercomparison,
Commission of European Communities
Medical Physics,
17(6), Nov/Dec 90, On the measurement of HVL in Film Screen Mammography,
L. Wagner, F. Cerra, B. Archer
Health Physics Journal,
Vol 37 (Dec), pp. 773-774, Evaluation of the MDH Industries, Inc, Pencil
Chamber for Direct Beam CT Measurements, William Pavlicek, John Horton,
Robert Turco
Measurement
of Radiation for X-Ray Computed Tomoghraphy Using Pencil-Shaped Ionization
Chambers, R.M. Gagne, T.B. Shope, J.L. McCrohan and C.K. Showalter
|
|
 |